MDS-3T Digital Ion-Selective Electrode (ISE) Sensor-MADSUR
Home > Water Quality Sensor > MDS-3T Digital Ion-Selective Electrode (ISE) Sensor
Products
  • Water Quality Sensor
  • Water Quality Testing
  • Monitoring instruments
  • Water quality monitoring station
  • Measuring equipment
  • Portable devices
  • Flow rate & velocity & level gauge
  • Laboratory testing equipment
  • Industrial process testing equipment
  • MDS-3T Digital Ion-Selective Electrode (ISE) Sensor

    The MDS-3T (RS485) is a multi-parameter sensor specifically designed for online measurement of ammonia nitrogen and nitrate nitrogen. The sensor uses the ion electrode method, so users only need to immerse the sensor in the water sample to directly read the ammonia nitrogen and nitrate nitrogen values. The sensor incorporates a built-in temperature sensor (Pt100) and two ion electrode interfaces (1 for ammonia nitrogen or nitrate nitrogen, and 2 for pH).

    E-mail:info@madsurgroup.com
    Send

    -Overview-

    The MDS-3T (RS485) is a multi-parameter sensor specifically designed for online measurement of ammonia nitrogen and nitrate nitrogen. The sensor uses the ion electrode method, so users only need to immerse the sensor in the water sample to directly read the ammonia nitrogen and nitrate nitrogen values. The sensor incorporates a built-in temperature sensor (Pt100) and two ion electrode interfaces (1 for ammonia nitrogen or nitrate nitrogen, and 2 for pH).


    -Benefits & Features-

    Economical and Reliable:

    - No sample pretreatment required; direct measurement of ammonia nitrogen and nitrate nitrogen.

    - pH measurement is standard.

    - Advantages: No reagents required, low operating cost.

     Flexible use and wide application range:

    - Wide measurement range:

    0.1-1000 mg/l NH4-N or 0.5-1000 mg/l NO3-N
    Simple and Safe Operation

    - Direct installation at the pool edge; no need for a collector or sample transfer pump.

    - Replace the membrane cap every 6 months to effectively extend electrode lifespan.

    - Standardized digital communication; plug and play.


    -Technical parameters-

    Measurement Range

    Ammonia nitrogen: 0.5-1000mg/L (NH4-N); Nitrous nitrogen: 1-1000mg/L (NO3-N); 

    Free fluoride: 0.5-10000 mg/L (F); Free chlorine: 2.5-10000 mg/L (Cl); Free calcium: 1-2000mg/L (Ca)

    Size

    Directly 32mm (maximum), length 166mm

    Working Life

    6 months (membrane), 3 months (electrolyte)

    Measurement Error

    Measurement value: ±10% (laboratory) or ±0.5mg/L

    Response Time

    < 2 min.

    Repeatability

    ±10% of the displayed value

    Environmental Temperature

    0-40℃

    Storage Temperature

    2-40℃

    Protection Level

    IP68

    Cable

    Standard 10-meter waterproof cable

    Instrument Power Consumption

    Within 6W





    WeChat MADSUR
    All rights reserved © 2026 Copyright MADSUR